Curetech Solutions
IBA Dosimetry

CC25 Ionization Chamber

Art. N°
Hover to zoom · click to enlarge

Request a quotation

We reply with pricing and availability for Art. N° .

Request a quote
Radiation Therapy

CC25 is intended for absolute and relative dosimetry of photon, electron, and proton beams in radiotherapy. Measurements can be performed in air, water phantoms, or solid phantoms.

CC25 is a high spatial resolution, small volume chamber. It is waterproof, vented through a waterproof sleeve, and fully guarded.

Brochure
/fileadmin/user_upload/products/02_radiation_therapy/CC04/Detector_Brochure.pdf
Inner dimensions
Inner dimensions - Active volume: 0.25 cm^3 - Total active length: 10 mm - Cylinder length: 7 mm - Inner diameter of outer electrode: 6.0 mm - Wall thickness: 0.4 mm (70 mg/cm2 ) - Diameter of inner electrode: 1.0 mm - Length of inner electrode: 7.5 mm
Used for Workflow
Detectors RT
Operational characteristics
Operational characteristics - Polarizing voltage: ±300 V (max. ±500 V) - Typical leakage current: 5 fA - Recommended pre-irradiation: 5 Gy - Typical sensitivity: 7.5 nC/Gy - Guard potential: ±300 V (max. ±500 V) - Temperature range: 15 °C – 35 °C - Relative humidity range: 20 % – 80 %